DIAGRAM A DIAGRAM B Diagram 1
Diagram 1 illustrates that even with 6 MV photons utilizing parallel opposed portals, if the treatment portal encompasses the posterior occiput or the frontal region, it is obvious that without wedges the target minimum dose is 10% lower than the prescribed central axis dose, and there is a hot spot in the thinner portion that can be 10% to 15% hotter. These differences can be minimized with the use of wedges as in Diagram 2. The tumor minimum is only 2% lower than the central axis dose, and the hot spots are smaller and of lower dose. Isodose distributions are required for parallel opposed fields. DIAGRAM 3 LARGE FIELD BOOST FIELD



Diagram 3 is an example of a composite plan for an anteriorly located lesion with significant edema. By combining large 15 wedged parallel opposed fields to 57.60 (Diagram 4), it was possible to even out hot spots and treat the target volume to a high dose throughout with minimal gradient. Optimization of individual plans and the submission of composite plans is an essential requirement of this study.


DIAGRAM 4

Diagram 5